# Angular-dependent resolution broadening parameters
#
# sigma_div_min_deg:
#   Residual angular broadening near the Bragg peak, in degrees of 2theta.
#   This should be very small for high-resolution measurements.
#   Suggested range for implanted/defective samples: 0.0002 - 0.002 deg.
#
# sigma_div_max_deg:
#   Effective angular broadening far from the Bragg peak, in degrees of 2theta.
#   This may include beam divergence, lateral incoherence, mosaicity, and implantation-induced disorder.
#   Suggested range for implanted/defective samples: 0.01 - 0.08 deg.
#
# p_div:
#   Exponent controlling how sharply the broadening increases away from the Bragg peak.
#   Low values make the broadening start closer to the peak.
#   High values delay the broadening until larger angular offsets.
#   Suggested range: 1.5 - 5.
#
# n_sigma_kernel:
#   Truncation range of the local Gaussian kernel, in units of sigma.
#   This is mainly numerical, not physical.
#   Suggested value: 4.
#
# Format:
# parameter_name = value

sigma_div_min_deg = 0.0005
sigma_div_max_deg = 0.055
p_div             = 1.5
n_sigma_kernel    = 4